Hach-Lange LDO Sensor User Manual

Page 3

Advertising
background image

Table of Contents

Specifications

on page 3

Operation

on page 9

General information

on page 4

Maintenance

on page 13

Product overview

on page 5

Troubleshooting

on page 15

Installation

on page 6

Replacement parts and accessories

on page 17

Calibration for measurements

on page 11

Specifications

Specifications are subject to change without notice.

Specification

Details

Wetted materials (probe body)

CPVC, sensor end and cable end

Polyurethane, over-molding on cable end and cable jacket

316 stainless steel body and screws

Viton, O-ring

Noryl, nut on the cable end

IP classification

IP68

Wetted materials (sensor cap)

Acrylic

Measurement range (dissolved oxygen)

0 to 20 ppm (0 to 20 mg/L)

0 to 200% saturation

Measurement accuracy (dissolved oxygen)

Below 5 ppm: ± 0.1 ppm

Above 5 ppm: ± 0.2 ppm

Repeatability (dissolved oxygen)

0.1 ppm (mg/L)

Response time (dissolved oxygen)

T

90

<40 seconds

T

95

<60 seconds

Resolution, sensor (dissolved oxygen)

0.01 ppm (mg/L); 0.1% saturation.

Measurement range (temperature)

0 to 50 °C (32 to 122 °F)

Measurement accuracy (temperature)

± 0.2 °C (± 0.36 °F)

Interferences

No interferences from the following: H

2

S, pH, K

+

, Na

+

, Mg

2+

, Ca

2+

,

NH

4

+

, Al

3+

, Pb

2+

, Cd

2+

, Zn

2+

, Cr (total), Fe

2+

, Fe

3+

, Mn

2+

, Cu

2+

,

Ni

2+

, Co

2+

, CN

, NO

3

, SO

4

2–

, S

2–

, PO

4

3–

, Cl

, Anion Active

Tensides, Crude Oils, Cl

2

< 4 ppm

Storage temperature

–20 to 70 °C (–4 to 158 °F)

Maximum ambient temperature

60 °C (140 °F). Suitable for use in water up to 50 °C (122 °F)

Hazardous location classification
(5790001 sensor only)

Class I Division 2, Groups A–D, T4 / Class I, Zone 2 Group 2C, T4

Note: This product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).

Certifications (5790001 sensor only)

ETL listed to ANSI/ISA, CSA and FM standards for use in
hazardous location.

Note: This product does not fulfill the requirements of the 94/9/EC Directive
(ATEX Directive).

Minimum flow rate

Not required

English 3

Advertising