4.5 CALIBRATE

Note: Turbidity measurement has been calibrated by the manufacturer — it does not need to be
calibrated again.

Note: It is imperative to calibrate for solid matter measurement (refer to section

4.5.2 Calibration of

the SOLID (TS) parameter

).

The zero point for turbidity and solid matter measurement has been set in the sensors by
the manufacturer.

Installation conditions in the pipes can cause interfering ground reflection when
measuring turbidity, which in turn may cause the zero point to shift. Compensate for this
effect with an offset correction (section

4.5.1.3 OFFSET

). If there are deviations between

the displayed measurement values and laboratory results that are unrelated to the factors
described above, the slope of the calibration curve can be adjusted using a factor (refer
to section

4.5.1 Calibration of the TURBIDITY (TRB) parameter

).

PROBE INFO

SENSOR NAME

Device name

EDITED NAME
INSTRUMENT
NUMBER
TURBIDITY

0.001 to 9999 FNU

SOLID

0.001 to 500 g/L

MODEL NUMBER

Sensor item number

CODE VERSION

Sensor software

COUNTER

MANUAL RESET PRESS ENTER: Security prompt

TEST / MAINT: COUNTER X DAYS BACKWARDS,

TOTAL: OPERATING HOURS COUNTER,

INTERVAL

Default for maintenance counter

SERVICE

SIGNALS

Explanation: refer to service manual

S5E1
S5E3
S6E1
S6E3
S5E2
S5E4
S6E2
S6E4

OUTPUT MODE

Behavior of instrument outputs in the SERVICE menu

HOLD
ACTIVE
TRANSFER
SELECTION

Service access

4.4.2 SENSOR setup

SELECT SENSOR (if there is more than one sensor)