Section 6 malfunctions, causes, resolution, 1 error messages, 2 warning messages – Hach-Lange TSS EX1 sc User Manual

Page 25: 1 error messages 6.2 warning messages, Refer to

Advertising
background image

25

Section 6

Malfunctions, causes, resolution

6.1 Error messages

Possible sensor error messages are displayed by the sc controller.

6.2 Warning messages

Possible sensor warning messages are displayed by the sc controller.

Table 1 Error messages

Error displayed

Definition

Resolution

MEAS OVERRANGE

Measurement range exceeded,
signals too small, probe can no longer
measure this concentration.

If error occurs more frequently: find another installation
location.

CAL. INSUFF. –

Calibration insufficient.

Probe requires another calibration point in a lower
concentration.

CAL. INSUFF. +

Calibration insufficient.

Probe requires another calibration point in a higher
concentration.

ZERO POINT

Calibration is

too close to the zero point.

Calibrate again with higher concentration.

CAL REQUIRED

No existing calibration

Calibrate probe.

EE RSRVD ERR

Error in the probe electronics

Call customer service department.

ERROR PROBE

Error in the probe electronics

Call customer service department.

LED FAILURE

Faulty LED

Call customer service department.

Table 2 Warnings

Warning displayed

Definition

Resolution

TEST / MAINT

Counter at zero

Call customer service department.