4 handling, 5 scope of delivery, 6 function check – Hach-Lange TSS EX1 sc User Manual

Page 9: 4 handling 2.5 scope of delivery

Advertising
background image

9

General information

Up to three calibration points can be defined for a strongly fluctuating medium. The
combined multiple-beam alternating light technique records solids in the medium with
even greater accuracy.

2.4 Handling

2.5 Scope of delivery

TSS EX1 sc sensor

Test log

Operating instructions

2.6 Function check

After unpacking and checking for any transport damage, perform a function check for all
components outside the ATEX zone.

1. Connect the sensor to the sc controller (refer to

3.3, page 12

).

2. Switch on the power supply to the sc controller.

The display is activated and the sensor enters measurement mode.

Note: The measurement value that is shown in air is not relevant.

3. If no warning or error messages are displayed, the function check is complete.

D A N G E R

Switch off the sc controller before connection or disconnection of the probe and sc controller.

N O T I C E

Protect the probe from impacts and never rest the probe on the measurement head.

N O T I C E

Never operate the probe within powerful magnetic fields! (EN50021 6.4)

Advertising