1 calibration of the turbidity (trb) parameter, 1 select turbidity (trb) parameter, 2 factor – Hach-Lange TSS EX1 sc User Manual

Page 19: 3 offset, 4 1 to 3 point calibration, If ther, Operation

Advertising
background image

19

Operation

19

Operation

19

Operation

At least a 1-point calibration must be carried out for a solid matter measurement. In
difficult application conditions, a 2-point or 3-point calibration may be necessary (refer to
section

4.5.2 Calibration of the SOLID (TS) parameter

).

4.5.1 Calibration of the TURBIDITY (TRB) parameter

Before the sensor can be calibrated to the TURBIDITY (TRB) parameter, the parameter
must be selected.

4.5.1.1 Select TURBIDITY (TRB) parameter

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Select CONFIGURE and confirm.

5. Press PARAMETERS and confirm.

6. Select the TRB parameter and confirm.

7. Go back to the MAIN MENU or the Measurement mode display.

4.5.1.2 FACTOR

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Press CALIBRATE and confirm.

5. Press FACTOR and confirm.

6. Set the desired factor and confirm.

7. Go back to the MAIN MENU or the Measurement mode display.

4.5.1.3 OFFSET

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Press CALIBRATE and confirm.

5. Press OFFSET and confirm.

6. Set the required offset and confirm.

7. Go back to the MAIN MENU or the Measurement mode display.

4.5.1.4 1 to 3 point calibration

Note: The turbidity measurement has been calibrated by the manufacturer.

Note: Before the sensor can be calibrated to the TRB parameter, the parameter must be selected
(refer to

4.5.1.1 Select TURBIDITY (TRB) parameter

).

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.