Section 5

Maintenance

The cleanliness of the measurement windows in the sensor head is decisive for the
accuracy of the measurement results!

Check the measurement windows for dirt and wear once a month.

5.1 Maintenance schedule

5.2 Clean the measurement windows

The windows are made of sapphire glass. The measurement windows can be cleaned
with any conventional cleaning agent and a soft cloth.

In the case of stubborn deposits, it is recommended that 5 % hydrochloric acid is used.

D A N G E R

The probe must not be opened. Work on the inside of the probe may only be carried out by the
manufacturer.

Maintenance task

Maintenance interval

Visual inspection

Monthly

Test calibration

Monthly (depending on environmental conditions)

Inspection

Every 6 months (counter)

D A N G E R

The probe (especially the type plate) must not come into contact with acetone or p-Xylene, or any
liquids containing these solvents.

D A N G E R

Potential danger with contact with chemical/biological substances.
Working with chemical samples, standards and reagents can be dangerous.
Make yourself familiar with the necessary safety procedures and the correct handling of the
chemicals before use and read and follow all relevant safety data sheets.

C A U T I O N

Observe safety conditions and wear protective clothing!
Safety glasses
Gloves
Overalls