2 calibration of the solid (ts) parameter, 1 select the solid (ts) parameter, 2 factor – Hach-Lange TSS EX1 sc User Manual

Page 20: 1 select the solid (ts) parameter 4.5.2.2 factor, 2 calibration of, The solid (ts) parameter, Operation

Advertising
background image

20

Operation

20

Operation

20

Operation

4. Press CALIBRATE and confirm.

5. Press CALIBRATE and confirm.

6. Press RECORD and confirm.

7. Press POINT... (point 1, 2 or 3) and confirm.

Once the calibration point has been recorded by the probe, a mark "<<" is displayed
after the point or points that has (have) been recorded for approximately 3 seconds.

Note: If the Calibrate menu is closed before the calibration is complete and then called again,
the "<<" mark is displayed again. This shows that the calibration for this point or points has not
yet been completed. The old calibration values are still being used.

8. Select the recorded POINT parameter and confirm.

9. Enter the laboratory comparison value and confirm.

To record more calibration points: repeat steps 6 to 9.

10. Go back to the MAIN MENU or the Measurement mode display.

The instrument automatically sorts the saved calibration points according to the size of
the calibration values, irrespective of the sequence in which the calibration points were
recorded.

Point 1 is always assigned to the smallest calibration value.

Point 2 is assigned to the next smallest calibration value.

Point 3 is assigned to the largest calibration value.

The value calculated in the laboratory can be corrected at any time by overwriting.

4.5.2 Calibration of the SOLID (TS) parameter

Before the probe can be calibrated to the SOLID (TS) parameter, the parameter must be
selected.

4.5.2.1 Select the SOLID (TS) parameter

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Select CONFIGURE and confirm.

5. Press PARAMETERS and confirm.

6. Select the parameter TS and confirm.

7. Go back to the MAIN MENU or the Measurement mode display.

4.5.2.2 FACTOR

1. Open the MAIN MENU.

2. Press SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Press CALIBRATE and confirm.

5. Press FACTOR and confirm.

6. Set the desired factor and confirm.

7. Go back to the MAIN MENU or the Measurement mode display.