Installation – Hach-Lange SONATAX sc User Manual

Page 16

Advertising
background image

16

Installation

If no measurement location without disruptive reflexes can be
found, select the probe menu SENSOR SETUP > CALIBRATE and
trigger a measurement in the PROFILE LIST menu item.

If there are increased values at the depth of the disruptive
reflexes in the PROFILE LIST which influence the
measurement value, another measurement location must be
selected. Sometimes it is sufficient to move the probe by a few
centimeters (refer to

Figure 9

, right).

If the reflections do not disrupt the measurement value, this can
be a suitable measurement location (refer to

Figure 9

, left).

3. Then assemble the tank edge attachment there and install the

probe.

Note: The probe should be installed

approximately 20 cm (7.9 in.)

deep

at the installation location (somewhat more than the entire probe head) to
make sure that it remains immersed during the entire measurement
operation.

4. In the CALIBRATE probe menu, enter the actual plunger depth

(Item A,

Figure 10

) in the PLUNGERDEPTH menu item.

5. Trigger after approximately 2 minutes another measurement of

the reflex list (to adjust the temperature sensor to the water
temperature) (also refer to

REFLEXLIST on page 24

).

6. In the CALIBRATE probe menu, enter the recorded value of the

tank depth (Item C,

Figure 10

) in the TANK DEPTH menu item.

Important note: The shown value of the tank depth is calculated
from information shown:

Tank depth C = plunger depth A + measured distance to floor B

This calculation is performed internally and leads to incorrect
results if the plunger depth is entered incorrectly.

Figure 9 Determine measurement location