Operation

KONFIGURIEREN (CONFIGURE)

EDIT NAME

Freely editable (up to 16 characters)

Factory setting: device number

PARAMETER

The measurement result can be shown as the sludge level (as the distance of the
sludge from the water surface) or as the sludge height (as the distance from the tank
floor). To calculate the sludge height, the tank depth specified in the TANK DEPTH
menu item is used.

(Sludge height = tank depth – sludge level)

Configurable: sludge level, sludge height

MEAS UNITS

Dimension of the measurement result.

Configurable: meters, feet

CLEAN. INTERVAL

Wipe interval,

Recommendation: 15 minutes

Configurable: 1 minute to 1 hour

RESPONSE TIME

Damping of the measurement value. In the case of high measurement value
fluctuations, a high damping of for example 300 seconds is recommended.

Configurable: 10 to 1800 seconds

LOGGER INTERVAL

The interval for the internal data log.

Configurable: 1, 2, 3, 4, 5, 6, 10, 15, 30 minutes

SET DEFAULTS

Reset to factory configuration for all menu items listed above. This occurs only after
a security prompt.

TEST / MAINT

PROBE INFO

SENSOR NAME

Display of the device name.

EDIT NAME

Display of the freely selectable measurement location

(factory setting: device number).

SERIAL NUMBER.

Device number

MODEL NUMBER

Item no. of the sensor.

HARDWARE-VERS

Production status of main circuit board

SOFTWARE-VERS

Sensor software version

COUNTER

WIPER COUNTER

Backward counter for wiping processes of the wiper profile.

After expiration of the counter, a warning message is shown. In the case of a wiper
change, the counter should be reset again.

TOTAL TIME

Operating hours counter

MOTOR

Forward counter for wiping processes.

TEST / MAINT

Date of the last performed maintenance.

REPLACE PROFILE

For the wiper profile change, the wiper arm moves to a central position. In this
position, the wiper arm can then be removed and installed without problems.




4.5 SENSOR SETUP menu (Continued)

Select the SONATAX sc (if several sensors are connected)