Section 1 technical data – Hach-Lange SONATAX sc User Manual

Page 5

Advertising
background image

5

Section 1

Technical data

Changes reserved!

General

Measurement method

Ultrasonic measurement (750–1250 kHz)

Measuring range

0.2 m–12 m (0.7 ft–40 ft) sludge level

Resolution

0.03 m (0.1 ft) sludge level

Accuracy

0.1 m (0.33 ft)

Response time

10–600 s (adjustable)

calibration

One time on commissioning

Environmental conditions

Ambient temperature

\> 0–50 °C (\> 0–122 °F)

Temperature compensation

Automatic

Flow speed

Max. 3 m/s

Pressure range

0.3 bar or

3 m (

43.55 psi or

10 ft)

Sensor specifications

Dimensions

130 mm

×

185 mm (5 in.

×

7.3 in.) (H

Ч

Ш)

Earth

Approximately 3.5 kg (123.5 oz) (without struts)

Maintenance requirements

< 1 hour/month, typically

Cable length

10 m (33 ft), maximum 100 m (330 ft) with extension cable

Power consumption

12 V, 2.4 W, (200 mA)

Protection type

IP68 (

1 bar (14.5 psi))

Compliance

CE, TÜV GS, UL/CSA

Materials

Probe body

Stainless steel 1.4581

Base plate and wiper

POM

Wiper magnet casting compound

Epoxy resin

Wiper rubber

Silicone rubber

Housing seals

NBR (acrylonitrile butadiene rubber)

Light guide seal

Polyurethane

Light guide

LEXAN polycarbonate

Sensor connection cable

(fixed connection)

1 cable pair AWG 22 / 12 VDC twisted, 1 cable pair AWG 24 / data twisted,
common cable shield, Semoflex (PUR)

Sensor connection plug (fixed
connection)

Type M12, protection type IP67

Cable gland

Stainless steel 1.4571

Cable gland insert

TPE-V

Cable gland O-ring

NBR, silicone

Advertising