Section 6 troubleshooting, 1 operating state led, 2 error messages – Hach-Lange SONATAX sc User Manual

Page 29: 1 operating state led 6.2 error messages, Section 6.1, Ssages; refer to

Advertising
background image

29

Section 6

Troubleshooting

6.1 Operating state LED

The sensor is equipped on the top side with an LED that provides
information on the operating state.

6.2 Error messages

In the case of an error, an error message is shown on the controller.
Find error messages and notes on the resolution of the errors in

Table 5

.

Figure 13 Operating state LED

1

Operating state LED

Table 4 Sensor status

Green LED

No errors or warnings

Green/red flashing LED

Probe in operation, probe position deviates strongly from the perpendicular, the
measurement value is retained, there is no error

Red LED

Error

LED off

No device function

Table 5 Error messages

Error message

Reason

Resolution

SENSOR MEASURE

The PROFILE LIST and graphic on the sc1000
is less than the programmed value for
LL THRESH.AUTO. over the entire tank depth,
or the ultrasound transducer is contaminated or
not immersed.

Check device data, also the plunger depth,
floor depth and advanced settings, check
installation and remove contamination.

POS. UNKNOWN

The wiper position is not detected. The wiper is
in the central position (after a wiper change).

Start a wiping process

Defective light barrier plate

Call service

Application with particles that block the wiper.

Cleaning of the ultrasound transducer and of
the wiper system.

AMPL DIAG

Internal error

Call service

MOIST

Moisture value \> 10

Call service

SENSOR ANGLE

The probe is more than 20° from perpendicular
and for longer than 180 seconds.

Check installation

Incorrectly calibrated position sensor

Call service

SYSTEM ERROR

Faulty RAM

Call service