Turbidity measurement, Measurement notes, Figure 6 – Hach-Lange 2100 N IS Laboratory Turbidimeter User Manual User Manual

Page 25: With a membrane

Advertising
background image

Figure 6 Prepare filtered sample using membrane or glass-fiber
filter

1 Filter pump

4 Stopper

7 Filter

2 Hose

5 Filter holder

3 Filter flask

6 Tweezers

Turbidity measurement

W A R N I N G

Potential explosion and fire hazard. This instrument is for measuring water based
samples. Do not measure solvent or combustible based samples.

For accurate turbidity readings use clean sample cells and remove air
bubbles. Refer to

Clean the sample cell

on page 16 and

Remove air

bubbles from the sample

on page 21.

Measurement notes

Proper measurement techniques are important in minimizing the effects
of instrument variation, stray light and air bubbles. For accurate and
repeatable measurements:
Instrument

• Make sure that the instrument is on a level, stationary surface that is

free of vibration during the measurement.

• Instrument stabilization is immediate. No warm-up time is necessary.
• Always close the sample compartment lid during measurement,

calibration and storage.

• Remove the sample cell from the instrument and turn off the

instrument if the instrument is stored for an extended time period
(more than a month).

• Keep the sample compartment lid closed to keep dust and dirt out.

Sample cells

• Always cap the sample cell to prevent spillage of the sample into the

instrument.

• Always use clean sample cells in good condition. Dirty, scratched or

damaged cells can result in readings that are not accurate.

• Make sure that cold samples do not “fog” the sample cell. Refer to

Prevent condensation on a sample cell

on page 22.

• Store sample cells filled with distilled or deionized water and cap

tightly.

• For the best accuracy, use a single sample cell for every

measurement or a flow cell.

Note: As an alternative, matched sample cells may be used for measurements but
do not provide as good of accuracy or precision as a single indexed sample cell or
flow cell. When using matched sample cells, align the orientation mark on the
sample cell with the reference mark on the sample cell holder.

Measurement

• Measure samples immediately to prevent temperature changes and

settling. Before a measurement is taken, always make sure that the
sample is homogeneous throughout.

• Avoid sample dilution when possible.

English 23

Advertising