Computer (rs232) commands, Advanced operation, Calibrate the turbidimeter with formazin standards – Hach-Lange 2100 N IS Laboratory Turbidimeter User Manual User Manual

Page 31: Prepare formazin standards

Advertising
background image

3. Push ENTER.
4. Use the arrow keys to change the value.
5. Push ENTER.
6. Push SETUP.

Computer (RS232) commands

A communication program (i.e., such as Window Terminal or ProComm
Plus) is recommended for computer operation of the instrument.
Configure the communication program to the RS232 connection settings.
Refer to

Configure the RS232 connection

on page 28.

Table 4

shows the RS232 command set for the instrument.

Table 4 RS232 command set

Command Description

VAL

Gets the current measurement with the measurement units.

LST

Gets the calibration standards and coefficients.

DAT

Gets the current date.
To change the date, enter DAT=MM/DD/YY (MM=month, DD=day,
YY=year), then push Enter.

TIM

Gets the current time in 24-hour format.
To change the time, enter TIM=HH:MM (HH=hour, MM=minutes),
then push Enter.

RMN

Gets the recorder minimum value.
To change the recorder minimum value, enter RMN=XXXXX
(XXXXX=a number, minimum value=0), then push Enter.

RMX

Gets the recorder maximum value.
To change the recorder maximum value, enter RMX=XXXXX
(XXXXX=a number, maximum value=10,000), then push Enter.

RTN

Gets the recorder trim minimum value.
To change the recorder minimum value, enter RTN=XXXXX
(XXXXX=a number, minimum value=200), then push Enter.

Table 4 RS232 command set (continued)

Command Description

RTX

Gets the recorder trim maximum value.
To change the recorder maximum value, enter RTX=XXXX
(XXXX=a number, maximum value=4800), then push Enter.

SAV

Gets the signal average buffer size.
To change the signal average buffer size, enter SAV=XX (XX=a
number, maximum value=15, default=10), then push Enter.

Advanced operation

Calibrate the turbidimeter with formazin standards

The instrument may be calibrated using prepared formazin standards
made from 4000-NTU formazin stock solution. Refer to

Accessories

on page 36.

Note: Use recently prepared formazin standards to get the accuracy specifications
for turbidity in

Specifications

on page 5.

Prepare formazin standards

For the best accuracy and long-term data comparability, use formazin
stock solution from Hach to make formazin standards.

Note: As an alternative, a 4000-NTU formazin stock solution that is prepared by
the user may be used to make formazin standards. Refer to

Making 4000-NTU

formazin stock solution

on page 32.

Prepare formazin standards immediately before calibration in an
environment that is at the same ambient temperature as the instrument.
Discard after use.
Refer to

Table 5

for the procedures to make the recommended

calibration standards.