Table 5 Formazin standard preparation

Standard

Step 1

Step 2

Step 3

20 NTU

Add 100 mL of
dilution water to a
clean 200-mL Class
A volumetric flask.
Refer to

Prepare

dilution water

on page 20.

With a TenSette

®

Pipet, add 1.00 mL of
well-mixed 4000-NTU
formazin stock
solution to the 200-
mL flask.

Dilute to the mark
with dilution water.
Stopper and mix.

200 NTU

Add 50 mL of dilution
water to a clean 100-
mL class A
volumetric flask.

With a TenSette

1

Pipet, add 5.00 mL of
well-mixed 4000-NTU
formazin stock
solution to the 100-
mL flask.

Dilute to the mark
with dilution water.
Stopper and mix.

1000 NTU Add 50 mL of dilution

water to a clean 100-
mL class A
volumetric flask.

With a TenSette

1

Pipet, add 25.00 mL
of well-mixed 4000-
NTU formazin stock
solution to the 100-
mL flask.

Dilute to the mark
with dilution water.
Stopper and mix.

1

A class A volumetric pipet may be used in place of a TenSette Pipet.

Calibration notes

• Make sure that the instrument is in the same ambient conditions as

where it is used.

• Make sure that the standards are at the same ambient temperature as

the instrument before use.

• Use only the provided silicone oil. This silicone oil has the same

refractive index as the vial glass and masks minor glass differences
and scratches.

• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and

the last calibration data is used. To exit a calibration and not save the
new values, push UNITS/Exit.

• In Calibration mode, automatic range and signal averaging on are

selected. When calibration is completed, all operational modes go
back to the last settings.

• All nephelometric (turbidity units of measure) calibrations are done at

the same time.

• The 4000-NTU standard does not have to be measured during

calibration if FNUs will be measured. Push CAL after the 1000 NTU
standard is measured to complete the calibration procedure.

• The FNU values of StablCal standards and formazin standards are

calculated using the conversion factors of 1 FNU = 1 NTU.

30 English