Flow cell maintenance, Flow cell operation, Flow cell storage – Hach-Lange 2100 N IS Laboratory Turbidimeter User Manual User Manual

Page 29: Using a manual flow cell, Use a cell adapter

Advertising
background image

6. Air dry the parts after cleaning.

Flow cell maintenance

• Keep all parts of the flow cell assembly clean.
• At intervals, replace all the tubing to make sure that the system is

clean. Keep the tubing as short as possible to minimize air locking and
lag time of sample flow. Locate the instrument as close to the drain as
possible.

Flow cell operation

• Do not use the flow cell for samples that contain large particles that

may collect and stop the sample from flowing.

• Slowly put the sample down the interior edge of the inlet reservoir to

prevent mixing of the sample, which can cause air bubbles. Air
bubbles create a false positive interference in a turbidity
measurement.

• If bubbles collect in the flow cell, gently tap the flow cell on a soft

surface to remove the bubbles. If bubbles continue to collect in the
flow cell, put the glass flow cell in liquid detergent for 24 hours and
then rinse fully.

• When measuring many samples of different turbidity, measure the

samples in order of the cleanest (lowest turbidity) to the dirtiest
(highest turbidity) to prevent contamination from one sample to the
next.

• Do not use greater than the recommended maximum sample pressure

of 34 kPa (5 psig).

• Keep the drain tubing below the center line of the instrument. If the

whole 152 cm (60 in.) length of drain tubing is used, make sure that
the end of the drain tubing is at least 46 cm (18 in.) below the center
line of the instrument.

• The flow-cell cover must be in place for the LED light source to

function.

Flow cell storage

• Install the reservoir cover when the system is not in use to prevent

contamination of the system by airborne particles.

• For short-term storage (a few hours), flush the system with distilled or

deionized water and leave the flow cell full of the flush water to
minimize air locks and build up of residue on the parts.

• For long-term storage, disassemble, fully clean and air dry all parts.

Using a manual flow cell
To set the flow rate, increase the height of the collection drain assembly
on the support rod to decrease the flow rate. Make sure that the bottom
of the collection drain assembly is no lower than 7.5 cm (3 in.) above the
support stand base.
To flush the flow cell, lower the collection drain assembly to the support
stand base to flush the flow cell.

Use a cell adapter

Many different test tubes, sample cells and ampules can be used to
measure samples when a cell adapter is used. Use a cell adapter when
the test tube, sample cell or ampule is less than 25 mm. Refer to

Accessories

on page 36 for the available cell adapters.

Use only test tubes and sample cells that are free of significant
scratches. Clean and apply silicone oil to all sample cells, test tubes and
ampules used with the cell adapters. Refer to

Clean the sample cell

on page 16.

Note: Performance specifications may be different than shown in

Specifications

on page 5 when test tubes, sample cells or ampules less than 25 mm are used.

Use a cell adapter when:

• Only a small quantity of sample is available.
• The sample to be measured is in an ampule that cannot be opened.

Refer to

Table 3

for minimum sample sizes.

Table 3 Minimum sample sizes

Test tube size

Sample

12 mm

2.5 mL

13 mm

3.5 mL

16 mm

5 mL

19 mm

7 mL

English 27

Advertising