Table 6 Error codes (continued)

Error

Description

Solution

ERR06 A/D is under the range

1. Make sure that no object is in the light

path.

2. Contact Customer Service if necessary.

ERR07 Light leak

1. Make sure that the cover for the sample

cell compartment is closed.

2. Turn the instrument off and then back on.

ERR09 Printer time out error

1. Make sure that the external printer is

connected correctly.

2. Make sure that the external printer is

selected (online).

ERR10 System voltage out of

range

1. Turn the instrument off and then back on.
2. Contact Customer Service if the error

occurs again.

ERR11 System loop test error

1. Turn the instrument off and then back on.
2. Contact Customer Service if the error

occurs again.

Diagnostic codes

Table 7

lists the diagnostic codes that are used to get information about

instrument operation when instrument operation is in doubt.
To do a diagnostic test:

1. Push SETUP.
2. Use the arrow keys to enter a diagnostic code.
3. Push ENTER to show the diagnostic value.
4. Push UNITS/Exit to go back to Measurement mode.

Note: To print a diagnostic report, hold down PRINT, then turn the instrument on.

Table 7 Diagnostic codes

Code

Display

Description

21

Pr In

Printer test

22

Test results are shown.

Display test

23

Test results are shown.

Keyboard test

24

Test results are shown.

Memory test

Delete calibration data

To delete any calibration data entered by the user:

1. Turn off the instrument.
2. Push and hold CAL.
3. Turn on the instrument.

The CAL? light flashes. The instrument starts in Calibration mode.

4. Calibrate the instrument before use.

Flashing 9s

When manual ranging is selected, the display will flash all 9s when the
sample being measured is greater than the selected range.
When automatic ranging is selected, the display will flash 9s when the
sample is greater than the maximum range of the instrument.Refer to

Measure over-range samples

on page 22.

Flashing 0s

When manual ranging is selected, the display will flash all 0s when the
sample measured is less than the selected range.
When automatic ranging is selected, the display will flash all 0s when the
measurement is less than the range of the instrument or a negative
value. Calibrate the instrument.

English 35