Using a flow cell, Install a flow cell, Clean a flow cell assembly – Hach-Lange 2100 N IS Laboratory Turbidimeter User Manual User Manual

Page 28: Install a flow cell clean a flow cell assembly, Figure 8, Show the methods for connecting the two types of

Advertising
background image

Figure 8 Standard shop air

1 Particle filter

5 Filter (Balston 100-12-BX or

equivalent)

2 Air dryer

6 Auto drain (Balston 20-105 or

equivalent)

3 Coalescing filter/regulator (0–

30 psig)

7 Filter housing (Balston

FR-920-30 or equivalent)

4 Shop air

Using a flow cell

C A U T I O N

Do not use a flow cell with flammable samples or those that contain
hydrocarbons, solvents, concentrated acids or concentrated bases that may
damage wetted parts of the cells. Conduct tests before use of flow cells if sample
compatibility is not known.

Note: Do not use a high pressure flow cell kit with this instrument.

Use a flow cell to increase the speed, accuracy and reproducibility of
measurement. The manufacturer especially recommends using a flow
cell for low turbidity measurements.

Install a flow cell

1. Fully clean and assemble the flow cell, tubing and stand. Refer to

Clean a flow cell assembly

on page 26 and the user instructions

provided with the flow cell.

2. Fill the flow cell and tubing with water and make sure that there are

no leaks or air bubbles.

Note: Air bubbles collect in areas that are not cleaned fully.

3. Clean the exterior surface of the flow cell with a soft, lint-free cloth to

remove water spots and fingerprints.

4. Apply a small bead of silicone oil from the top to the bottom of the

flow cell.

Note: Use only the provided silicone oil. This silicone oil has the same
refractive index as the flow cell glass and masks minor glass scratches.

5. Use the oiling cloth provided to apply the oil equally to the surface of

the flow cell. Remove the excess oil. Make sure that the flow cell is
almost dry.

Note: Put the oiling cloth in a plastic storage bag to keep the cloth clean.

6. Install the flow cell in the sample cell compartment.
7. Push the inlet and outlet tubes in the slots on the top of the

instrument enclosure so the sample cell cover can be installed. Refer
to the user instructions.

8. Put the flow-cell light cover over the flow cell.

Note: The standard sample cell cover of the instrument does not close when
the flow cell is installed.

Clean a flow cell assembly

1. Disassemble the flow cell assembly.
2. Clean the inside and outside of the glass parts with a laboratory

glass cleaning detergent. Follow with multiple rinses with distilled or
demineralized water.

Note: All tubing, flow cells, and caps in the flow cell assembly can also be
steam sterilized.

3. If measuring low turbidity samples, clean the inside and outside of

the glass parts with 1:1 hydrochloric acid and rinse multiple times
with dilution water.

4. Fill the sample cell with distilled or demineralized water and

immediately put the caps on the sample cell.

5. Clean the inside and outside of the plastic parts and tubing with

laboratory detergent and warm water.

Note: At intervals, replace the tubing as contaminants, including
microbiological growths, are difficult to remove from the inside surface of the
tubing.

26 English