Formazin calibration procedure – Hach-Lange 2100 N IS Laboratory Turbidimeter User Manual User Manual

Page 33

Advertising
background image

Formazin calibration procedure

For the best accuracy, use four matched sample cells or the same sample cell for all measurements during calibration. Refer to

Matching sample cells

on page 19.

1. Push CAL.
The S0 light turns on.
The NTU value of the
dilution water used in
the last calibration is
shown.

2. Rinse a clean
sample cell two times
with dilution water. Fill
the sample cell to the
line (about 30 mL) with
dilution water and
immediately put the cap
on the sample cell. Use
the same dilution water
that was used to
prepare the formazin
standards.

3. Clean the sample
cell with a soft, lint-free
cloth to remove water
spots and fingerprints.
Do not invert the
sample cell.

4. Apply a small bead
of silicone oil from the
top to the bottom of the
sample cell.

5. Use the oiling cloth
provided to apply the oil
equally to the surface of
the sample cell.
Remove the excess oil.
Make sure that the
sample cell is almost
dry.

6. Put the sample cell
in the sample cell
holder with the triangle
on the sample cell
aligned with the
reference mark on the
sample cell holder.
Close the cover.

English 31