Maintenance, Clean the meter, Apply silicone oil to a sample cell – Hach-Lange 2100Q and 2100Qis Basic User Manual User Manual

Page 11

Advertising
background image

Option

Description

Signal
Average

The Signal Average mode compensates for reading
fluctuations caused by drifting of sample particles through
the light path.
The X-bar icon is shown on the display when signal
averaging is on.
The Signal Average mode measures 12 times and starts to
show the average after three readings. The final result is
the average of all 12 readings.

Rapidly
Settling
Turbidity

(RST)

The Rapidly Settling Turbidity (RST) mode calculates and
continuously updates the turbidity reading of the sample to
a confidence of 95%, based on the accumulated trend of
the real time measured values.
The RST mode is best used on samples that settle rapidly
and continuously change in value. The reading is based on
a correctly prepared sample that is homogeneous at the
beginning of the reading. It is best applied to samples that
are greater than 20 NTU. The sample must be mixed
thoroughly by inversion immediately before inserting it into
the meter.
The target icon is shown on the display when the Rapidly
Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates five
readings while showing intermediate results.

Maintenance

C A U T I O N

Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.

Clean the meter

The meter is designed to be maintenance-free and does not require
regular cleaning for normal operation. Exterior surfaces of the meter may
be cleaned as necessary.

Note: Do not clean the meter with solvents to avoid damaging the material.

1. Clean the meter with a dust- and lint-free dry or slightly damp cloth. A

mild soap solution can also be used for liposoluble contamination.

Apply silicone oil to a sample cell

Sample cells and caps must be extremely clean and free from significant
scratches. Apply a thin coating of silicone oil on the outside of the
sample cells to mask minor imperfections and scratches that may
contribute to light scattering.

Note: Use only the provided silicone oil. This silicone oil has the same refractive
index as the sample cell glass.

1. Clean the inside and
outside of the cells and
caps by washing with a
laboratory glass
cleaning detergent.
Follow with multiple
rinses with distilled or
demineralized water.

2. Apply a small bead
of silicone oil from the
top to the bottom of the
cell.

3. Use the provided
oiling cloth to spread
the oil uniformly. Wipe
off the excess so that
only a thin coat of oil is
left. Make sure that the
sample cell is almost
dry with little or no
visible oil.
Note: Store the oiling
cloth in a plastic
storage bag to keep the
cloth clean.

English 11