Laurel Electronics Custom ASCII Protocol For Laureate Series 2 Digital Panel Meters, Counters, Timers & L-Series Transmitters User Manual
Page 12

12
6. FRONT PANEL SETUP, SERIAL COMMUNICATIONS
6.1 FRONT PANEL SETUP, DIGITAL PANEL METERS & SCALE METER ONLY
Press Menu
Select Key
Press Digit Select
Key
Press Value Select Key
__000
Output filtering
0
Send unfiltered signal
1
Send filtered signal
__000U
Baud rate
0
300 baud
1
600 baud
2
1200 baud
3
2400 baud
4
4800 baud
5
9600 baud
6
19200 baud
.
SEr 1
.
Press
until
Ser 1
is displayed.
Fixed Parameters:
- No parity
- 8 data bits
- 1 stop bit
__000U
Output update rate,
Continuous Data Output
Mode.
60 Hz
50 Hz
0
0.017 sec
0.020 sec
1
0.28 sec
0.34 sec
2
0.57 sec
0.68 sec
3
1.1 sec
1.4 sec
4
2.3 sec
2.7 sec
5
4.5 sec
5.4 sec
6
9.1 sec
10.9 sec
7
18.1 sec
21.8 sec
8
36.6 sec
43.5 sec
9
72.5 sec
86.7 sec
_0000U
Line feed
0
No <LF> following <CR>
1
<LF> following <CR>
_0000U
Alarm data with readings
0
No alarm data with reading
1
Alarm data with reading
_0000U
Control of data output
0
Continuous data output
1
Data output on ASCII command only
.
SEr 2
.
Serial Setup 2
_0000U
Meter address
Select
1
thru
F
for addresses 1 thru 15.
Select
0.
thru
F.
(with decimal point) for
addresses 16 thru 31.