Laurel Electronics Custom ASCII Protocol For Laureate Series 2 Digital Panel Meters, Counters, Timers & L-Series Transmitters User Manual
Page 51

51
1 1 Relay1 Off when Al1 active, Relay2 Off when Al2 active
0B (U)
Alarm Confg2
Bit 7 6 5 4 3 2 1 0 Alarm Trigger Delay
60 Hz
50Hz
0 0 0
0.018s
0.021s
0 0 1
0.035
0.043
0 1 0
0.07
0.085
0 1 1
0.14
0.17
1 0 0
0.28
0.34
1 0 1
0.56
0.68
1 1 0
1.13
1.36
1 1 1
2.27
2.72
Alarm Comparison
0 0
Al1 Net, Al2 Net
0 1
Al1 Gross, Al2 Net
1 0
Al1 Net, Al2 Gross
1 1
Al1 Gross, Al2 Gross
0D (U)
Alarm Cnfg 4
Same as for DPM, page 35
0C (U)
Alarm Cnfg 3
Same as for DPM, page 35
BF (L)
Analog Output
Setup
Bit 7 6 5 4 3 2 1 0
0 0 Filtered Net
0 1 Filtered Gross
1 0 Unfiltered Net
1 1 Unfiltered Gross
0 0
Current Output (0-20 mA)
0 1
Voltage Output (0-10V)
1 0
Current Output (4-20 mA)
1 1
Voltage Output (-10 to +10V)
31 (L)
Serial Cnfg1
Bit 7 6 5 4 3 2 1 0 Continuous Output Data Rate
60 Hz
50 Hz
0 0 0 0
0.017s
0.020s
0 0 0 1
0.28
0.34
0 0 1 0
0.57
0.68
0 0 1 1
1.1
1.4
0 1 0 0
2.3
2.7
0 1 0 1
4.5
5.4
0 1 1 0
9.1
10.9
0 1 1 1
18.1
21.8
1 0 0 0
36.3
43.5
1 0 0 1
1:13
1:27