Insufficient bead count, Failed control in batch, Failed curve fit – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 153

Advertising
background image

x

MAP Technology

Troubleshooting

PN 89-00002-00-061 Rev. A

9 - 7

Insufficient Bead Count

•

Do not select Cancel during sample acquisition. Selecting the
Cancel function while a sample is acquiring causes an error
message to occur.

•

Ensure that you selected the correct template.

•

If applicable, see if all of the tests (beads) were added to the
samples.

•

See if the samples were properly mixed before they were
dispensed into the wells.

•

See if you loaded the sample to the correct well on the plate.

•

Check if the sample volume added to the well is correct.

•

Ensure that the samples were completely resuspended prior to
running on the analyzer.

•

Remove and sonicate or flush the XYP instrument Sample Probe
to remove possible clogs.

•

Check if the XYP instrument Sample Probe is properly aligned.

•

Complete the following analyzer functions from the Maintenance
tab: Drain, Backflush, and Alcohol Flush.

•

Run the system control microspheres.

•

Check to see if the delta cal temperature < 3.0 degrees.
Recalibrate if greater and retry.

•

If the system controls pass, contact the kit manufacturer.

Temperature Divergence
from Calibration
Temperature

•

Calibrate and verify the system.

•

If this message still appears following calibration, contact
Luminex Customer Support.

We recommend that you install the system in a temperature-
controlled environment.

Failed Control in Batch

This message occurs when one or more of the assay controls are not
within the expected range.

•

Ensure that you selected the correct template.

•

Ensure that the correct control lot number and target values were
selected and entered correctly.

•

Verify that the system is functioning properly by running the
system controls.

•

If the system controls pass, contact the kit manufacturer.

Failed Curve Fit

This message occurs when the calculation could not occur for the
assay standards for data interpretation.

•

Ensure that you selected the correct template.