Set the XYP

Instrument Heater

Temperature

Refer to your assay kit instructions to see if the assay needs to be
analyzed at a particular temperature. If the instructions indicate that
the XYP instrument heater is needed, you must manually set the
heater to the specified heat setting. Use the heater only with the XYP
instrument heater block in place.

Any temperature that you set remains in effect until you set another
temperature or turn off the XYP instrument plate heater.

The system displays the target temperature in the entry field below
the ON button. Before the heater block temperature reaches the new
temperature setting, the XYP Heater Temperature thermometer
appears red. Upon reaching the target temperature, the thermometer
turns green. See “System Monitor” on page 7-31 for more
information about the system monitor.

Prior to being turned on or set, the ON button appears gray and
disabled.

‹

To set the XYP instrument heater temperature:

1. Insert the XYP instrument heater block into the plate holder.

2. Click the ON button. The light on the button turns green and the

thermometer fluid turns red as it raises to reach the target
temperature.

3. Click and drag the blue arrow on the XYP instrument Heater

Temperature thermometer to the temperature you want the XYP
instrument heater block to maintain.

4. Wait for the heater to reach your selected temperature and

stabilize, and the thermometer to turn green before processing
samples.

Command List

The Command List lists commands associated with batches or multi-
batches that you have loaded for processing on the system. The
command list also tells you the status of each command as the
system begins processing it. For instance, the command list shows
that the system is washing while processing a wash command. The

Note:

The heating function is

not intended for use with
standard 96-well microtiter
plates.

Warning:

The heater plate of the XYP instrument is hot when in

use and may cause personal injury. Do not touch the heater
plate.