Accuracy and precision, Sensitivity, Capacity – Luminex 100 IS User Manual, Version 2.1 User Manual

Page 42

Advertising
background image

Luminex 100 IS User Manual Version 2.1

x

MAP Technology

5 - 4

PN 89-00002-00-061 Rev. A

Accuracy and Precision

•

Sample uptake volume ± 5%

•

Classification of microspheres > 80%

•

Misclassification of microspheres < 0.5%

•

Temperature control 0°C to +2°C of target

•

Internal sample carry over < 0.9%

•

Soluble background fluorescence emission at 575 nm
automatically subtracted from fluorescence intensity values

Sensitivity

•

Detect 1000 fluorochromes phycoerythrin (PE) per microsphere

•

Reporter channel dynamic range: 3.5 decades

Capacity

The specifications below reflect minimum capacity values:

•

10 GB hard drive

•

Store data for up to 240,000 test results

•

1.4 MB 3½” floppy disk

•

100 MB read/write CD ROM

•

Analyze multiple 96-well plates per batch

•

Analyze multiple assay templates per plate

•

Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample

•

Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1-100 unique xMAP microspheres sets
in a single sample

•

Reporter channel range: 3.5 decades of detection

•

Sample core: 15 - 20 µm core at 1 µL/sec sample inject rate

•

Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)

•

Automatic sampling from a 96-well plate

•

Start sampling from any well position

•

Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills

•

Microtiter plates with 96 wells must be compatible with the XYP
instrument plate holder. The following microtiter plate types are
compatible with the XYP instrument plate holder: flat bottom,
conical, round, filter bottom, half plates [overall height no more
than 0.75” (19 mm)], any color

•

Microtiter plates with 96 wells must be compatible with XYP
instrument heater block temperature from 35°C to 55°C (95°F to
131°F) when performing heated assays and using the heater
block.