100 IS S Users Manual Version 2.1

x

MAP Technology

7 - 26

PN 89-00002-00-061 Rev. A

Run Self

Diagnostics

Run self diagnostics to see if the device and all system operations are
running correctly. The system tests these functions when running self
diagnostics:

•

Flash memory test

•

Microcontroller RAM test

•

Nonvolatile memory test

•

DSP program CRC test

•

DSP capture test

•

High voltage module

•

Channel backgrounds test

•

Pressurization test

•

Actuator test

•

Syringe pump test

•

Backflush valve test

‹

To run self diagnostics:

1. Click the Self-Diagnostics button from the maintenance tab. A

confirmation dialog box appears.

2. Click OK. The system begins processing the various self

diagnostic tests. As it completes the tests, the status bar changes
from a “Processing” command state to an “Idle” command state.

The self-diagnostic test should take less than one minute to
complete. If the self diagnosis failed, the system lets you obtain
more detailed information regarding the results of the self-
diagnostic test.