6.0  PARAMETERS

 

 

The balance has 7 parameters that can be set by the user. 

FUNCTION 

SECTION 

DESCRIPTION 

F1 UNT

See section 6.1 

Sets the units to be used 

g

 / 

ct / Lb / oz / d / GN / OZt / dWt / MM / TL.T / TL.C / TL.t /t 

/ N /g2. 

F2 EL

See section 6.2 

Sets the backlight  

EL on

:  backlight always on 

EL  AU

:  backlight  automatically  turns  on  when  a  key  is  
pressed 

EL oFF

: backlight always off 

F3 SEr

See section 6.3 

Sets the print parameters  

F4 oFF

See section 6.4 

Sets the auto power‐off parameter  

F5 IEC

See section 6.5 

Internal or external calibration select  

F6 CA

See section 6.6 

Cal mass fine adjust  

tECH

 

Technical parameters setting mode / factory setting 

6.1 

ENABLING

 

WEIGHING

 

UNITS 

You can enable and disable the weighing units available to the user when they press 
the [Mode] key as described in section 4.5 Weighing Units.    

1) To  set  this  parameter  press  the  [Mode]  key  during  self‐checking  test  when 

turning on the balance. 

2) After a few seconds, the display will show the first function F1 UNT.  

3) Press the [Tare] key to view the current settings of each unit.  

4) Pressing  the  [Tare]  key  will  cycle  through  the  other  units  along  with  their 

current settings. For example, if oFF is displayed with the weighing unit Carats, 
the user will not be able to use this unit while weighing. 

5) Pressing  the  [Mode]  key  will  change  the  setting  of  a  particular  unit.  For 

example,  to  enable  the  weighing  unit  Carats,  change  the  setting  to  on  by 
pressing the [Mode] key. 

14 | 

P a g e  

 

 

 

 

 

 

© Adam Equipment Company 2008

 

Advertising