7.0  CALIBRATION 

The Highland series of balances comes standard with HandiCal

TM

 internal calibration 

to make calibrating the balance quick and easy. However you can also calibrate the 
balance using an external verification weight if needed. The HandiCal method is the 
default but if you would like to use external calibration then you must first enable 
this via the parameter (section 6.5 Selecting The Internal Or External Calibration).  

 

Internal Calibration using HandiCal  

1) Press the [on/off] key to turn the power on.  

2) Press  the  [Smpl]  and  [Print]  keys  at  the  same  time  during  the  self‐checking 

test. 

3) The display will show unload. Remove any weight from the stainless steel pan.  

4) When the stable indicator shows press the [Tare] key.  

5) The display shows C Int. Press the [Tare

]

 key 

6) The  display  shows  LoAd.  Lower  the  internal  calibration  mass  located  behind 

the pan. Use the handle to help lower the weight as far as it will go. Press the 
[Tare] key once the stable sign is shown.  

7) The  display  will  show  PASS.  Rotate  the  weight  back  to  its  off  position.  Once 

removed the balance will return to zero.   

 

External Calibration 

1) Press the [on/off] key to turn the power on.  

2) Press  the  [Smpl]  and  [Print]  keys  at  the  same  time  during  the  self‐checking 

test. 

3) The display will show unload. Remove any weight from the stainless steel pan.  

4) When the stable indicator shows press the [Tare] key.  