Nova-Tech IHS 2a User Manual

Page 21

Advertising
background image

5) The  display will show the first weight that  you can use to calibrate  the  unit. 

You can change this value by pressing the [Mode] key. The weights that can 
be used are as follows:  

Model # 

HCB123  HCB153 

HCB302 

HCB602  HCB602H

HCB1002  HCB1502  HCB3001 

Weight 1 

60g 

50g 

100g 

200g 

200g 

500g 

500g 

1000g 

Weight 2 

120g 

100g 

200g 

400g 

400g 

1000g 

1000g 

2000g 

Weight 3 

‐ 

150g 

300g 

600g 

600g 

‐ 

1500g 

3000g 

 

6) Once the calibration weight has been selected press the [Tare] key 

8) The display shows LoAd. Place your weight on the pan. Press [Tare] key once 

the stable sign is shown.  

9)  The display will show PASS. Remove the weight from the pan.  

 

NOTE: If the calibration fails retry.  The balance will show FAlL H (when the weight is 
higher)  or  FAlL L  (when  the  weight  is  lower).  Repeat  the  process  using  the  correct 
calibration weight. 

 

21 | 

P a g e  

 

 

 

 

 

 

© Adam Equipment Company 2008

 

Advertising
This manual is related to the following products: