4.0  BASIC

 

OPERATION 

4.1 

TURNING

 

ON

 

THE

 

BALANCE 

Plug in the unit using the AC adapter or use the internal rechargeable battery. It is 
recommended that you charge the battery for at least 8 hours before first use.  

1) To turn on press the [on/off] key once and release. The balance will show the 

software revision and the capacity then self‐test before showing zero on the 
display and the stable sign. 

2) The balance is ready to be used.  

3) To turn the balance off after use press the [on/off] key again. There is an auto 

power‐off  function  that  will  automatically  turn  the  unit  off  if  not  used  for  a 
period of time, this can be set in the parameters section. 

If  an  error  is  shown  while  turning  on  the  balance  check  the  internal  calibration 
weight has not been left in the calibrate position. Move the lever all the way to the 
left.   Turn off the balance and turn it back on again. 

4.2 

ZEROING

 

/

 

TARE 

You can press the [Tare] key to set a new zero point and show the zero reading if 
the  weight  reading  is  less  than  4%  of  the  total  of  the  maximum  capacity  of  the 
balance.  This  may  be  necessary  if  the  weight  reading  is  not  reading  zero  with 
nothing on the pan. The zero indicator will show up in the upper left corner of the 
LCD.  

If  you  are  using  a  container  to  weigh  then  you  can  place  this  on  the  platform  and 
press  the  [Tare]  key.  Providing  the  container  weight  is  more  than  4%  of  the 
maximum capacity of the balance, the digits will show zero and NET will light up on 
the  display.  You  can  then  weigh  your  sample  in  the  container.  Taring  weight 
subtracts from the total balance capacity.  

 

Note: When the container is removed a negative value will be shown 
equivalent  to  the  total  value  of  the  amount  tared.  The  balance  will 
not tare a value for a container unless the stability light indicates that 

the weight is stable thus ensuring a correct taring function. 

9 | 

P a g e  

 

 

 

 

 

 

© Adam Equipment Company 2008

 