Nova-Tech pH-CON 700 User Manual

Page 10

Advertising
background image

5

e buffer group

See Section 9.6

to change th

1)

Press

as needed to select pH.

Dip the pH and ATC electrodes into pH buffer and press

2)

. The

secondary display will lock on the appropriate buffer value. Provide

ing for best results. When the READY indicator appears, pr

stirr

ess

to accept. The primary reading will flash briefly before the

secondary display begins scrolling the remaining available buffers.

Rinse the pH and ATC electrodes then dip into the next pH buffe
The secondary display will lock on the appro ate buffer value.

When the READY indicator appears, press

3)

r.

pri

to accept. The

primary reading will flash briefly then display the percent efficiency
(slope) before the secondary display begins scrolling the remaining

4)

available buffers.

To calibrate another buffer repeat step 3) or press

to return to

the measurement mode. Note: The meter will automatically return
to measurement mode upon successful completion of the numbe
specified calibration points. To spec

r of

ify a different number of pH

int (offset) calibration is only allowed with pH 7.00 or

pted during a new calibration,

ca

calibration points see Section 9.6.

IPS:

A single po

T
pH 6.86 buffers.

When the first calibration value is acce
all prior

libration values are erased.

ress

P

at any time to abort calibration and return to measurement

USA buffer group

1.68, 4.01, 7.00, 10.01, 12.45

NI

5

ST buffer group

1.68, 4.01, 6.86, 9.18, 12.4

mode.