Calibration error.
Buffer value does not
match displayed value or
electrode is disconnected
or failing.

Use fresh buffer solutions.
Check electrode connection.
Clean & recondition electrode.
Replace electrode.

Broken or worn electrode

Replace electrode

Unstable pH
reading

External ‘noises’ or
induction (electrical ‘noise’
caused by a nearby motor)

Remove or switch off interfering
device

Slow
response

Dirty electrode.

Clean & rehydrate the electrode

CAL

during pH
CAL mode

Calibration was attempted
using only 1-point that was
not pH 7 (USA) or pH 6.86
(NIST).

Recalibrate using 2 or more pH
points or use pH 7 (USA)
or use pH 6.86 (NIST).