5 manual conductivity & tds calibration procedure, 6 temperature calibration, Manual conductivity & tds calibration procedure – Nova-Tech pH-CON 700 User Manual

Page 14: Temperature calibration

Advertising
background image

4) When

the

READY

indicator appears, press

to accept. The

t

primary reading will flash briefly before returning to measuremen
mode upon successful calibration.


5.5 Manual Conductivity & TDS Calibration Procedure

1) Press as needed to select conductivity (μS or mS) or TDS

(ppm or ppt) calibration.

2) Dip the electrode into the calibration standard and press

.

Provide stirring for best results.
The primary display will show the current reading,

3)

while the

secondary display will be the factory default value. Adjust the

9

conductivity or TDS reading using

or

. Press

to

accept. The primary reading will flash briefly before returning to
measurement mode upon successful calibration.

5.6 Temperature Calibration
The thermistor sensor used for automatic temperature compensation

req

mended upon probe

nt,


1)

n with a known accurate temperature

e bath.

2) Press

and measurement is accurate and stable, so frequent calibration isn’t

uired. Temperature calibration is recom

replaceme

whenever the temperature reading is suspect, or if

matching against a certified thermometer is desired.

Place the probe into a solutio
such as a constant temperatur

followed by

. The primary display shows the

) Adjust the temperature using

measured temperature while the secondary display shows the
factory default temperature.

4

or

. Press

to accept or

n adjustable maximum value of ± 5

º

C

to cancel. The meter allows a

o

º

( r ± 9 F) from the factory default temperature.