8 3.4 tds (tds factor)—con & tds only, 4 tds (tds factor)—con & tds only, 5 t.co (temperature coefficient)—con & tds only – Nova-Tech pH-CON 700 User Manual

Page 19

Advertising
background image

Press

or

to select “Yes” or “No”.

Press

to confirm.

9.8 3.4

n & TDS only

ress


tdS (TDS factor)—Co

P

.

Press

or

to select the desired TDS factor (.40 to

1.00).

Press

to confirm.

9.9 3.5 t.CO (Temp

efficient)—Con & TDS only

The tem

c

e amount of change in conductivity per

his will

erature Co

erature oefficient is th

p

degree temperature (% per ºC). The PC 700 is factory set to a
temperature coefficient of 2.1 % per ºC. For most applications t
provide good results. The meter allows adjustment from 0.0 to 10.0.

TIP:

Select 0.0% for uncompensated measurements. The temperature

will be measured by the electrode and displayed in measurement
mode—without compensation.

See Section 11 – Calculating Temperature Coefficients.

Press

.

Press

or

to select the desired value (0.00 to 10.0).

Press

to

confirm.

9.10 3.6 t.nr (Normalization Temperature in ºC)—Con & TDS only
When Au

ompensation is used, measurements

are adjusted by the temperature coefficient to the normalization
temperature. The default value is 25 ºC.

tomatic Temperature C

14