9.

SETUP FUNCTIONS

Use the setup feature to customize your instrument operation. First,
select the appropriate measurement mode you wish to adjust by

pressing

until the desired units are displayed (i.e. pH, mV,

Conductivity, or TDS). During measurement, press and hold

for 5

seconds to enter SETUP mode of the parameter being measured.

Press

or

to change programs or change options.

Press

to select the program or confirm selection.

Press

to go back one level or return to measurement mode.

9.1 1.0 CAL (Calibration)

Press

to view stored calibration points.

9.2 2.0 ELE (Electrode Information)

Press

to view mV offset (OFS)—mV and pH modes.

Press

to view % slope (SLP)—pH mode.

Press

to view cell efficiency in each range—conductivity

and TDS modes.


TIP:

When multiple pH slopes exist, the slope that is available to view

will depend on the measured value before entering SETUP mode.
For example, if pH calibration values pH 4.01, 7.00 and 10.01 were
completed, there will be two slope values; one that corresponds to pH
values below 7.00 and another that corresponds to pH values above
7.00. If the measured pH value was 5.23 when setup mode was
entered, the pH slope that corresponds to values lower than 7.00 will

12

be available to view. If no calibration is stored, the offset will be 0.0
mV and the slope is 100%.