4.2 Millivolt (mV) Offset Adjustment
Oxidization Reduction Potential (ORP or Redox) is not a precise
measurement, but is useful as a relative indicator. As such, mV offset
adjustment is not meant to enhance accuracy, but rather to make
readings comparable to a reference. Commercial ORP solutions are
often used as a check s
verified to be close to a given value, instea of being used as a
calibration standard in which adjustments

made in an attempt to

1)

Connect an ORP electrode and press

as needed to select

(or R.mV).

mV

2)

Dip the ORP electrode into a solution with a known mV value
(i.e. Zobel, Light’s, quinhydrone, or iodidetriiodide) and stir.

3)

Press

when the reading is stable

he prima

display shows

the relative millivolt value (R.m ) whi

he secon ry display sho

factory default mV value.

Adjust the R.mV value using

. T

ry

V

le t

da

ws

the

)

or

. Press

4

to accept or

to cancel. The meter allows an adjustable maximum value of

150

±

mV from the factory default mV value.

es

mV.


Note:

When an offset has been stored successfully, R.mV replac

5.

CONDUCTIVITY AND TDS CALIBRATION

5.1 Automatic or Manual Calibration
The PC 700 is capable of automatic or manual calibration for
conductivity, and manual calibration for TDS.

In the automatic calibration mode, the meter will automatically se
one of (4) conductivity calibration standard values depending on the
range and normalization temperature being used (TABLE 1).