8

Date & Time

Setting the correct date and time is required for GLP and will apply to power off,
measurement, data log, and print functions. Instrument has battery backup to
retain date/time settings upon power loss. Factory reset will not apply to date and
time setting once it has been set. Changes related to daylight savings time must
be manually entered.

Date Format: Select (MM DD YY) or (DD MM YY)

Time Format: Select 12Hrs (AM/PM) or 24Hrs

Password

Select “ENABLE” to restrict access to Calibration and Setup modes. When
password protection is enabled, password entry is required before performing any
calibration, or making changes to the setup mode. Setup parameters can be
viewed, but can not be changed without correct password entry. The password is
a user selectable number from 1 to 99999.

Select “DISABLE” if password protection is not desired.

The meter does not allow you to edit setup parameters or perform a new
calibration unless you enter the correct password. If an incorrect password is
entered 3 consecutive times, the meter returns to measurement mode.

In the event the password can not be recovered, a password can be provided via
a written request to Eutech Instruments/Oakton Instruments. The instrument
serial number and your contact information are required.

Clear Datalog

Select “YES” to delete all saved data that was manually or automatically stored
into memory. This step is most commonly required when the “TIMED” data log
interval is not turned off and as a result, fills the memory to capacity. When the
memory is full, deleting the saved data is required in order to save additional
data/ The 2700 will never erase old data in favor of new data when the memory is
full.

If you wish to view, print or send the saved data to a PC before deleting, use the
MEM OUT” function from measurement mode.