Additional dissolved oxygen calibration notes:

• Keep the membrane free from contact with solid objects.

• Use with aqueous solutions only.

• Do not submerse the probe past the immersion limit (see below).

See

Section 22—DO Electrode Maintenance

for additional information.

• Calibration adjustment is limited to ±40% of the factory default value to

prevent erroneous calibrations.

• If % calibration is attempted with a value of 10.1% to 49.9%, a

“Calibration Error” will result. The following table lists corresponding
calibration values in % saturation calibration mode:

% Saturation

(of factory default value)

Calibration Value

less than 10%

0%

10.1% to 49.9%

Calibration Error

50% to 150%

100%

• Calibration of the concentration mode will only replace the previous

concentration calibration and does not affect the % saturation calibration.

To offset your % saturation reading to match another instrument, see

Section 15—Offset Calibration

.

• Perform daily calibration for best results. New calibration values will

automatically override the previous calibration.