Nova-Tech 2700 Series Benchtop Meters User Manual

Page 17

Advertising
background image

13







4) The primary display is the un-calibrated measured value. The 2700

automatically searches for and selects the appropriate value from your buffer
group in the secondary display. This value will blink when the ‘Stable’
indicator appears. Selecting CLR-C will clear the existing calibration.


5) Press ENTER to accept the calibration value

of the measured buffer. When successful the
accepted value will be visible inside of a
beaker icon. Percentage slope is visible after
two or more calibration points have been
completed.




6) Repeat Steps 4 & 5 with additional pH buffers

in any order or press ESC to escape and save
your calibration. When the specified number of
calibration points is met, the pH calibration
report page is automatically displayed.

To specify a different number of pH calibration
points see

Section 4—Cal Points

.

If the meter is password protected, you will be prompted to
enter a

password. See

Section 3—Password

.