0% calibration, Mg/l or ppm (concentration) calibration – Nova-Tech 2700 Series Benchtop Meters User Manual

Page 39

Advertising
background image

35

0% Calibration
7) After completing steps 1 thru 6 for 100% saturation calibration, place your

electrode in zero oxygen solution (0.08M sodium sulfite). Continuous stirring
is not required.


8) The primary reading is the current measured

value. Be patient—the reading can take
several minutes to reach 0% saturation!


9) When the reading has finally stabilized and

locked on 0% press ENTER to confirm the
calibration.



mg/L or ppm (Concentration) Calibration
1) Connect the DO electrode and power on DO 2700 for 5-15 minute warm up.

2) Press MODE if needed for concentration measurement mode.

3) Rinse the probe well with clean water then place into your solution with a

known dissolved oxygen concentration value (i.e. determined by titration or
another instrument).

4) Press CAL and switch the stirrer on. The upper reading is the current

measured value. The lower reading is the factory default reading for
comparison, along with salinity and temperature correction shown.

A manual DO concentration calibration

5) Allow the reading to stabilize. Use the up/down arrows to adjust the upper

reading to match the DO concentration of your solution.

The minimum

concentration calibration value is 2.00 mg/L. Press ENTER to confirm the
calibration.

Advertising