Conductivity, TDS, Salinity, & Resistivity Setup

Use Setup mode to customize Conductivity, TDS, Salinity, & Resistivity
parameters.

Press

from the measurement screen and then press ENTER when

Conductivity, TDS, Salinity, or Resistivity is selected.

Sample ID

This is a user selectable number from 1 to 99999.
Incorporating a sample ID to identify one or more
data points is useful to distinguish data that is
saved into memory or sent to a PC or printer. Use
the up/down arrow keys to adjust the values and
left/right arrow keys to move the cursor.


Cal Method

(conductivity mode only)

The 2700 is capable of automatic or manual
calibration for conductivity, and manual calibration
for TDS, Salinity, and Resistivity. This option
allows you to select AUTO (automatic) or
MANUAL conductivity calibration.

In the automatic calibration mode, the meter will
automatically select one of (4) conductivity
calibration standard values depending on the range and normalization
temperature being used (see table below).

Automatic Calibration Values

Normalization Temperature

Range

#

Conductivity

Range

25 ºC

20 ºC

r 1

0.00 – 20.00 µS

None

None

r 2

20.1 – 200.0 µS

84 µS

76 µS

r 3

201 – 2000 µS

1413 µS

1278 µS

r 4

2.01 – 20.00 mS

12.88 mS

11.67 mS

r 5

20.1 – 200.0 mS

111.8 mS

102.1 mS