Experiment 2: standing waves in a tube, Introduction, Procedure – PASCO WA-9612 RESONANCE TUBE User Manual

Page 13: Equipment needed, Figure 2.1 equipment setup, Manual auto, On off on off

Advertising
background image

012-03541E

Resonance Tube

9

Experiment 2: Standing Waves in a Tube

EQUIPMENT NEEDED:

— PASCO Resonance Tube
— Function Generator
—Frequency Counter (if your function generator does not accurately indicate frequency)
— Oscilloscope (recommended, but not necessary)

Introduction

A sound wave propagating down a tube is reflected back and forth from each end of the tube,
and all the waves, the original and the reflections, interfere with each other. If the length of the
tube and the wavelength of the sound wave are such that all of the waves that are moving in the
same direction are in phase with each other, a standing wave pattern is formed. This is known as
a resonance mode for the tube and the frequencies at which resonance occurs are called resonant
frequencies. In this experiment, you will set up standing waves inside the Resonance Tube and
use the miniature microphone to determine the characteristics of the standing waves.

Procedure

➀ Set up the Resonance Tube, oscilloscope, and function generator as shown in Figure 2.1. Turn on the

oscilloscope. Set the sweep speed to 5 ms/div and the gain on channel one to approximately 5 mV/div.

Figure 2.1 Equipment Setup

T T L

H I

G N D

L O

M I N

R A N G E

A D J U S T

M A X

O U T P U T

F R E Q U E N C Y

A M P L I T U D E

P I - 9 5 8 7 B
D I G I T

A L F U N C T I O N

G E N E R A

T O R - A M P L I F I E R

H E RT Z

W AV E F O R M

I N P U T

G N D

E X T E R N A L

BK PRECISION

200 Mhz OSCILLISCOPE

MODEL 2120

INTENSITY

FOCUS

TRACE NOTATION

TRIG LEVEL

COUPLE

SOURCE

SLOPE

λ

- Y

TIME/DIV

X-POS

VAR

VAR

VAR SWEEP

CAL

CAL

mV

V

CH 1 VOLTZ/DIV

CH 2 VOLTZ/DIV

CAL

mV

V

VERTICAL MODE

PULL XS

PULL XS

CH 2

CH 1

AC

DC

AC

DC

AC

CH1

CH2

ALT

EXT

POS

POS

NORM
EXT

CH1
CH2

NORM
EXT

CH1
CH2

MANUAL AUTO

T X-Y

T X-Y

LINE

CAL

EXT CH4

POWER

200V MAX

400V MAX

400V MAX

-

+

+

-

1

2

3

4

5

SPEAKER INPUT

.1 W MAX

1

2

3

4

5

ON

OFF

ON

OFF

Oscilloscope

Function generator

Microphone and

probe rod

Amplifier

Turn on the amplifier and the function generator. Set the output frequency of the function generator
to approximately 100 Hz. Adjust the amplitude of the function generator until you can distinctly
hear the sound from the speaker. If you use the oscilloscope, trigger on the speaker output.

WARNING: You can damage the speaker by overdriving it. The sound from the speaker

should be clearly audible, but not loud. Note also that many signal generators become more
efficient and thus produce a larger output as the frequency increases, so you may need to
reduce the amplitude as you increase the frequency.

Advertising