Appendix A: Quad Pressure Calibration

The purpose of this calibration is to fine-tune all four pressure measurements so they
read the same when exposed to the atmosphere. This will allow the small pressure dif-
ferences that occur in the apparatus to be measured more accurately.

Conduct this procedure with all four pressure ports exposed to the same pressure.

DataStudio

1.

Click the Setup button to open the Experiment Setup win-
dow.

2.

Click the Calibrate Sensors button to open the calibration
window (see Figure 3).

3.

At the top of the Calibrate Sensors window, select Quad
Pressure Sensor
.

4.

Select the Calibrate all similar measurements simulta-
neously
option.

5.

Select the 1 Point (Adjust Offset Only) option.

6.

Click Read From Sensor (in the Calibration Point 1 section
of the window).

7.

Click OK.

Xplorer GLX (Standalone Mode)

1.

Press

+

to open the Sensors Screen.

2.

Press

again to open the Sensors menu.

3.

From the menu, select Calibrate to open the Calibrate Sensors win-
dow (see Figure 4).

4.

In the first box of the window, select Quad Pressure Sensor.

5.

In the third box of the window, select Calibrate All Similar Mea-
surements
.

6.

In the Calibration Type box, select 1 Point Offset.

7.

Press

(Read Pt 1).

8.

Press

(OK).

Figure 3: DataStudio calibration window

F4

F4

F3

F2

F1

Figure 4: GLX calibration window

F4

F3

F1