Teledyne 6750 - Total Organic Carbon analyzers User Manual

Page 19

Advertising
background image

Total Organic Carbon Analyzer

Theory of Operation

Teledyne Analytical Instruments

7

(sparges) the CO

2

converted from the TIC and it is measured by an

independent CO

2

detector as “TIC”. POC is also stripped from the

sample in the sparger as HC, but the HC is undetected by the CO

2

detector and does not interfere with the TIC measurement.

A portion of the same sample is also directed to the oxidation

reactor, where inorganic and all organic carbon is converted to CO

2

and

detected by a second CO

2

detector as “Total Carbon”.

A mathematical subtraction of the TIC from the TC measurement

yields a complete, Total Organic Carbon analysis performed as a “TOC-
True” method.

TOC - True = TC - TIC


The difference between the Model 6750’s TOC-True Method

and other simple (TC-TIC) subtraction methods is that most other
simple TC-TIC subtraction methods only deal with a TC measurement,
then a separate TIC measurement and a simple mathematical difference
to derive a “TOC” analysis.

In the TOC-True method used by the Model 6750, the analyzer

actually performs both TC and TIC measurements simultaneously and
continuously on the same sample. The TOC Measurement Basics and
Methods are illustrated in Figures 2-3 and 2-4.
In conclusion:

1. NPOC is the preferred method if no POC/VOC is present

in the sample.

2. TOC-True is the preferred method if POC/VOC exists in

the sample and a TOC analysis is desired which includes
all organic carbon species, for a TOTAL Organic Carbon
measurement.




Advertising