Total Organic Carbon Analyzer

Maintenance

Teledyne Analytical Instruments

45

5.3 Module Service

5.3.1 NDIR (Gas Calibration)

1. From the Menu, select [CALIBRATE].

The following screen will appear:

2. Select [INFRARED ANALYZER]. The following

Screen will appear:

3. Disconnect tubing from reactor output and connect IR

calibration adapter (P/N ST20026) as shown in Figures 5-
1, 5-2, and 5-3.

4. Connect facility tubing to “Zero” and “Span” Gas Bottles