Total Organic Carbon Analyzer

Maintenance

Teledyne Analytical Instruments

65

6. Remove 4 pump mounting screws

for pump removal.

7. Replace motor and reassemble in

reverse step sequence.

Figure 5-14: Removing the Pump

Note: Actual mounting panel may appear somewhat different but

mounting hardware is the same.

5.3.12.2 P

UMP

H

EAD

T

UBING

R

EPLACEMENT

1. Separate the end bells (the pump head

halves).

2. Hold the end bell containing the rotor as

shown with the tubing retainer grooves
pointing down.