Total Organic Carbon Analyzer

Installation

Teledyne Analytical Instruments

25

3.3.3 Alarm Connections

The Model 6750 is equipped with two concentration alarms and

one analyzer malfunction alarm. The three alarm-circuit connectors use
spring terminals for making connections to internal alarm relay
connects. Each alarm provides a set of Form C contact with both
normally open and normally closed contact connections. The contact
connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 amperes at 250 VAC into a resistive load.
See “As Built Drawings” in the Appendix. The connections are:

• Threshold Alarms:

• Can be configured as high (actuates when concentration

is above setpoint), or low (actuates when concentration is
below threshold).

• Can be configured as failsafe or non-failsafe.

• Can be configured as latching or non-latching

• Malfunction Alarm

3.3.4 RS-232 Serial Connection

An RS-232 port is provided for serial communication (MODBUS).

Connect a standard RS-232 cable from the computer to the RS-232 port
on the main board.

3.4 Gas/liquid Connections

Refer to the “As Built Drawings” in the Appendix for details and

Figure 3-5.

The Model 6750 Analyzer provides easy access for liquid, gas, vent

and drain connections.

3.4.1 Liquid Connections

Liquid phase sample connections include the following: (See

Figure 3-5):

• Sample (On-Line)

• D. I./Dilution Water

• Calibration Standard