Theory of Operation

Model 6750

Teledyne Analytical Instruments

12

actual measurement of the TIC, a more accurate TOC is achieved since
(TOC = TC - TIC).

“TOC-Direct” mode is preferred for accuracy when no or little

volatiles are present. In this mode, the carbonate-free sample is
extracted from the organic carbon reactor and measured directly as
TOC. Pump (P –1) continuously introduces the carbonate-free sample
to the reactor. No valves or complicated injection mechanisms are
required. The remaining organic carbon is oxidized to CO

2

in the

reactor. The resultant gaseous stream is directed to a gas/liquid
separator and then to the NDIR. The CO

2

gas is analyzed by the

infrared analyzer (NDIR) as a direct correlation of Total Organic Carbon
(TOC).

A computer controls all functions and outputs. A “Touch Screen”

provides operator interface and a paperless chart recorder utility.

Figure 2-6: Sample Handling for TOC -Direct