Fairbanks NexWeigh User Manual
Page 48

Section 5: Serial I/O
04/14
48
51216 -- Rev. 6
5.4.8. NexWeigh Serial Commands and Definitions,
Continued
f)
Gross mode
Definition: Used to switch scale to gross mode
Command:
‘G’
Options:
none
Sample Command: "G \r"
g)
Gross or Net mode
Definition: Used to toggle scale between gross and net mode
Command:
‘GN’
Options:
none
Sample Command: "GN \r"
h)
Set AZT band
Definition: Used to set the Auto zero tracking band. Only valid settings allowed.
Command:
‘SAT’ ‘divisions’
Options:
0.5, 1, 3, 0
Sample Command: "SAT 0.5\r"
Sample Command: "SAT 3\r"
i)
Set Scale Id
Definition: Used to set the scale’s id field. Only valid ids between MINSCALEID and
MAXSCALEID allowed.
Command:
‘SSI’ ‘ID’
Options:
1-32
Sample Command: "SSI 25\r"
j)
Set Output Mode
Definition: Used to set the serial output mode on the RS232 port. Polled, Polled w/id,
Continuous, Printer, Auto Printer, Remote display.
Command:
‘som’ ‘mode’
Options:
POLL, POLLID, CONT, PRTR, APRT, REMD
Sample Command: "som POLL\r"