Logger setting – General Tools and Instruments LTH8809DL User Manual

Page 6

Advertising
background image

LOGGER SETTING

5

There are two methods to set logger
parameters: Through PC or via meter's
keypad.

Press "SET" key to enter set mode.
There are 6 parameters in set mode:
-Sample Point:
'PtS' & 'P1' show on the screen .
-Start Mode:
' ' & 'P2' show on the screen.
-Start Time:
'Stt' & 'P3' show on the screen.
-Sample Rate:
'Sr' & 'P4' show on the screen.
-Alarm Set:
'AL' & 'P5' show on the screen.
-Real time:
'rtC' & 'P6' show on the screen

Press "Up" /" Down" to choose the mode.
Press "SET" to enter each mode.
Press "MODE" to escape to previous
status.

Sample Point:
- After pressing "SET " key to enter, the
selectable sample point values will
appear. They are 1, 2, 4, 8, 12, 16.
1 means 1000 sample points. ....16
stands for 16000 points. For the models
measure relative humidity, 1 means
500 Temp +500 RH points , 16 means
8,000 Temp +8000 RH points.

By Keypad Setting

NOTE: If the start mode is not
Scheduled, you won't see the Start
Time in the setting menu
.