2.2.10 Serial commands


The HP Series are automatically sending data through serial port at 9.5 Hz (push mode).
There is no need to ask for readings.

The HP sends the following data:

Data

ID

Units

Power

Pw=

Watts

Water Temperature

Tin=

Celsius

Water Flow

F=

Liters per minute


Example) Pw= 0.0 Ti= 20.000 F= 0.000


HP communication parameters:

a.

Bits per second: 57600

b.

Data bits : 8

c.

Parity: None

d.

Stop bits: 1

e.

Flow control: None



Serial Commands List:

The star (*) is part of each command and the commands are case sensitive.

Command

Description

Applies to

*RST

Hardware Reset

The HP will return to default settings.

USB and

DB-15 connector

*VER

Returns information about HP type and firmware version

Example:

HP__A-_KW-H_, Version X.XX.XX

USB only

*F01

Returns information about HP

Serial number, calibration wavelength (nm), DB15 sensitivity

(mV/W)

Example:

SerialNumber: 123456, Lambda : 1064, Sensitivity: 0.X00000

USB only

*F02

Returns information about HP current status

Current wavelength (nm), Thermal offset (W), Offset factor (W),

Multiplication factor and Anticipation status (On/Off).

Example:

Wavelenght: 1064, Thermal Offset : 0, Offset : 0, Multiplier : 1,

Anticipation: On

USB only

*PWC

To set the Wavelength

Wavelength correction command (5+ characters).

Example :

*PWC00808 to selects the wavelength 808 nm

USB only

*OFF

Sets the thermal (zero) offset

USB only